Smeg Tethys D60 User Manual page 55

Washer disinfector
Table of Contents

Advertisement

ALARM ID
DESCRIPTION
Water being taken into chamber at
AF:26
wrong time. Cold water intake
valve "EVF" may have failed.
Demineralised water being taken
into chamber at wrong time. Cold
AF:28
water intake valve "EVF" may have
failed.
Chamber does not empty. No
AF:29
drainage.
During the working cycle, the
AF:30
chamber water level exceeds the
safety level. Water safety level.
Water standing in washing
AF:32
chamber with appliance in
standby.
Steam condenser water intake
AF:33
anomaly. No water in steam
condenser.
Steam condenser water drainage
AF:34
failure.
User Manual
Check the water supply:
Check that the water supply pressure is within the permitted
range.
Follow the DEFAULT PROCEDURE described above. If the alarm
does not disappear, follow the RESET PROCEDURE.
Also check that there are no leaks from the appliance (no water
around it).
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
Check the water supply:
Check that the demineralised water supply pressure is within the
permitted range.
Follow the DEFAULT PROCEDURE described above. If the alarm
does not disappear, follow the RESET PROCEDURE.
Also check that there are no leaks from the appliance (no water
around it).
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
Check the connection to the water drain,
especially that the height of the drain connections complies with
the specified requirements and that there are no restrictions in
the drain hoses.
Check the water supply to the appliance:
1. Intake pressure.
2. That the connections are correct, as specified in this manual.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
Indicates that there is no water in the steam condenser when
there should be.
Check the water supply:
1. Intake water pressure
2. That the connections are correct, as specified in this manual.
3. Blockages or restrictions in hoses.
Follow the DEFAULT PROCEDURE described above. If the alarm
does not disappear, follow the RESET PROCEDURE.
Check that the drain connections are correct:
1. Height and position of the drain connection
2. That the connections are correct, as specified in this manual.
3. Blockages or restrictions in hoses.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
Tethys D60 - Tethys T60 - Tethys T45
USER ACTION
Page 55 - 77

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tethys t60Tethys t45

Table of Contents