Max voltage/power
There are three possible
electrical versions.
60cm models: The names of the
models have different suffixes
to identify the power supply
rating.
E.g. If the model does not have
a suffix, e.g. "Tethys D60" it will
be three-phase, with neutral,
400V between phases.
"Tethys D60-3" indicates three-
phase model, 230V between
phases, no neutral.
"Tethys D60-1" indicates single-
phase model.
DATA COMMUNICATION
- PRINTER CONNECTION
MAX. NOISE LEVEL
USE
ALTITUDE
AMBIENT TEMPERATURE
MAX RELATIVE HUMIDITY
INSTALLATION CATEGORY (OR SURGE CATEGORY or
INSTALLATION CATEGORY)
ELECTRICAL INSULATION CLASS (ref. IEC 61140)
DEGREE OF POLLUTION
CONFORMITY
MEDICAL DEVICE
CLASSIFICATION
User Manual
ELECTRICITY SUPPLY
MODEL
Tethys D60
Tethys T60
Tethys D60-3
Tethys T60-3
Tethys D60-1
Tethys T60-1
Tethys T45
OTHER DATA
AMBIENT CONDITIONS
STANDARDS AND CLASSIFICATION
- European directive 2014/35/EU (2006/95/CE) [Safety]: European Standard EN61010-
1:2010, EN61010-2-040:2015
- European directive 2014/30/EU (2004/108/CE) [Emc] : European Standard
EN61326:2013
- European directive 2011/65/CE (RoHS 2) : European Standard EN50581:2012
- European directive 93/42/CE, 2007/47/CE [MD]: European Standard
EN ISO 14971:2012, EN 62304:2006+AC:2008, EN62366:2008, EN ISO 15883-
1:2009+A1:2014, EN ISO 15883-2:2009
- Other: European Standard EN 61770:2009
class IIb medical devices (in accordance with the classification criteria established by
Directive 93/42 and subsequent additions and amendments, extended and amended by
EC Directive 2007/47).
Tethys D60 - Tethys T60 - Tethys T45
ELECTRICITY SUPPLY
400V 3N~ / 50Hz / 12A / 7000W
230V 3~ / 50Hz / 19A / 7000W
230V 1N~ / 50Hz / 12A / 2800W
230V 1N~ / 50Hz / 14A / 3300W
RS-232 PORT
50dB
Indoor
Up to 1000m
From 5°C to 40°C
80% for temperatures up to 31°C with linear decrease
to 50% at the temperature of 40°C
2 (rif. IEC61010-1:2010, par.3.6.8)
CHARACTERISTICS
RECOMMENDED
MAGNETOTHERMAL
SWITCH
[IN THE SYSTEM, RESPONSIBILITY
OF THE USER]
FUSES 10.3x38 16 A ON
POWER SUPPLY
[L1, L2, L3]
FUSES 10.3x38 20 A ON
POWER SUPPLY
[L1, L2, L3]
FUSES 10.3x38 16 A ON
POWER SUPPLY
[L1, N]
FUSES 10.3x38 16 A ON
POWER SUPPLY
[L1, N]
II
I