10 INSTALLATION

The device's characteristics and positioning and installation instructions are provided below.
10.1

TECHNICAL DATA

The documentation includes a description of the rated characteristics of the electrical connection, all the intake and
outlet connections and the ambient conditions for which the appliance was designed.
For "non electric" intakes, the pressure and flow ranges are provided (ref. point 5.4.2 IEC61010-2-040:2015).
Type of water required
Pressure
Type of threaded connection
Required flow [min - max]
Mains water hardness [max]
Fe2+ / Fe3+ [max]
Cold water temperature [max]
Demineralised water conductivity [max]
Limits depending on microbial
contamination
Series
Standard External [Height x Width x
Depth]
Standard External - aquastop accessory
[Height x Width x Depth] - overall height is
increased by +7mm
Net weight
Weight of appliance + packaging
Washing chamber
External surface
User Manual
WATER SUPPLY REQUIRED
1 - Mains (required)
(for models equipped with a steam condenser, the use of purified water with hardness
Minimum microbiological quality required: "potable water" (ref. limits
in accordance with Italian Legislative Decree 31/2001)
DIMENSIONS and WEIGHT
Tethys D60, Tethys T60
850 x 600 x 600
857 x 600 x 600
72 kg
90 kg
MATERIALS
Tethys D60 - Tethys T60 - Tethys T45
200 kPa – 500 kPa (2 bar – 5 bar)
3 / 4 "
2 - 12 litres / min
42 °f
below 15°f is recommended).
0.5 ppm
35°C
30μS/cm
AISI 316L
AISI 304
2 - Demineralised
(recommended)
Tethys T45
850 x 450 x 620
-
56 kg
66 kg