Smeg Tethys D60 User Manual page 75

Washer disinfector
Table of Contents

Advertisement

WARNING
The drain must comply with international regulations: our company accepts no liability for
pollution caused by improper use of the appliance.
General guidelines for installing the drain
The use of a drain with syphon trap is recommended. During installation, the following precautions should be
complied with:
 Since the drain water temperature is 95°C, the end of the drain hose must be permanently connected to the hose
connector, using the ties provided.
 There must be no sharp bends in the drain hose which might cause restrictions.
 The difference in height between the end of the drain hose and the surface on which the appliance is installed
must comply with the stated specifications.
 The end of the hose must never be submerged in water.
 The inside diameter of the drain pipe to which the hose is connected must be at least 40 mm.
 A drain pipe 50 mm in diameter is recommended.
 No extensions must be added to the drain hose supplied with the appliance. Any extensions might cause problems
of flow-back into the chamber.
User Manual
fig. 48 – Adapter with connector for 1/2" hose.
Tethys D60 - Tethys T60 - Tethys T45
Page 75 - 77