1.
Assemble the Model G Laboratory Stand (refer to assembly instructions in Appendix I)
2. Attach the Viscometer head to the clamp on the laboratory stand rod.
3. Connect the RTD probe to the socket on the rear panel of the DVPlus.
4. Remove the shipping cap which secures the coupling nut on the Viscometer to the pivot cup.
5. Make sure that the power switch at the rear of the DVPlus is in the OFF position. Connect the power cord
to the power supply provided. Do not use any other power supply or the warranty may be voided. Plug the
power supply cable into the DVPlus power in outlet only. Important, if the power supply cable is plugged in
anywhere else damage may occur and warranty may be voided
6. Turn the power switch to the ON position and allow the Viscometer to warm up for 10 minutes before
leveling and performing autozero.
7.
If the DVPlus prompts that the Viscometer must be leveled, follow the instructions on the screen to level
the instrument. Please note: The leveling prompt will not appear if instrument has already been leveled.
8. If appropriate, connect USB cable (DVP-202) to USB port for connection of DVPlus to the PC.
Temperature Probe

1.6 Safety Symbols and Precautions

Safety Symbols:
The following explains the safety symbols that may be found in this operating manual.
Indicates hazardous voltages may be present.
M21-400 REV.1.1
The DVPlus must be used with the provided power supply.
The power supply must be earth grounded to maintain compliance with electrical certifications.
D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L
Figure 1-5
USB B
USB A
Power in