Ametek BROOKFIELD DVPlus Operating Instructions Manual page 50

Digital viscometer
Table of Contents

Advertisement

Vane Spindles
Spindle
V-71
V-72
V-73
V-74
V-75
V-71
V-72
V-73
V-74
V-75
V-71
V-72
V-73
V-74
V-75
V-71
V-72
V-73
V-74
V-75
V-71
V-72
V-73
V-74
V-75
Note:
1 Pa = 10 dyne/cm
Viscosity Range is given at a rotational speed of 10 RPM
HB is the highest torque range available
M = 1 million
K = 1 thousand
Pa = Pascal
Special Considerations
In taking viscosity measurements with the DVPlus Viscometer, there are two considerations that pertain to the low
viscosity limit of effective measurement:
1.
Viscosity measurements should be accepted within the equivalent % Torque Range from 10% to
100% for any combination of spindle/speed rotation.
2.
Viscosity measurements should be taken under laminar flow conditions, not under turbulent flow
conditions.
M21-400 REV.1.1
Torque Range
NOT RECOMMENDED FOR USE ON LV TORQUE
LV
LV
LV
LV
RV
RV
RV
RV
RV
HA
HA
HA
HA
HA
HB
HB
HB
HB
HB
5xHB
5xHB
5xHB
5xHB
5xHB
2
cP = centiPoise
mPa•s = milliPascal•seconds
D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L
Shear Stress Range
(Pa)
0.188-1.88
0.938-9.38
9.38-93.8
3.75-37.5
0.5-5
2-20
10-100
100-1K
0-400
1-10
4-40
20-200
200-2K
80-800
4-40
16-160
80-800
800-8K
320-3.2K
20-200
80-800
400-4000
4K-40K
1.6K-16K
Viscosity Range
cP(mPa•s)
104.04-1.04K
502-5.02K
5.09K-50.9K
1.996K-19.96K
262-2.62K
1.11K-11.1K
5.35K-53.5K
54.3K-543K
21.3K-213K
524-5.24K
2.22K-22.2K
10.7K-107K
108.6K-1.086M
42.6K-426K
2.096K-20.96K
8.88K-88.8K
42.8K-428K
434.4K-4.344M
170.4K-1.704M
10.48K-104.8K
44.4K-444K
214K-2.14M
2.172M-21.72M
852K-8.52M
50

Advertisement

Table of Contents
loading

Table of Contents