Appendix E - Calibration Procedures - Ametek BROOKFIELD DVPlus Operating Instructions Manual

Digital viscometer
Table of Contents

Advertisement

APPENDIX E - CALIBRATION PROCEDURES

The accuracy of the DVPlus is verified using viscosity standard fluids which are available from AMETEK Brookfield or
your local authorized dealer. Viscosity standards are Newtonian, and therefore, have the same viscosity regardless of
spindle speed (or shear rate). Viscosity standards, calibrated at 25°C, are shown in Table F-1 (Silicone Oils) and Table
F-2 (Mineral Oils).
Container Size:
Temperature
Conditions:
Normal 25
Viscosity (cP)
5
10
50
100
**25ºC, 93.3ºC, 149ºC, Refer to AMETEK Brookfield catalog for more information
M21-400 REV.1.1
For Viscosity Standards < 30,000 cP, use a 600 mL Low Form Griffin Beaker having a working
volume of 500 mL.
For Viscosity Standards > 30,000 cP, use the fluid container. Inside Diameter: 3.25" (8.25 cm)
Height: 4.75" (12.1 cm)
Note: Container may be larger but may not be smaller.
As stated on the fluid standard label: (+/-) 0.1°C
The DVPlus should be set according to the operating instructions. The water bath must
be stabilized at test temperature. Viscometers with the letters "LV" or "RV" in the model
designation must have the guard leg attached (see Appendix F for more information on the
guard leg).
C
0
Mineral Oil Viscosity Standard Fluids
BEL Part No.
B29
B200
B600
B1060
B2000
B10200
B21000
B73000
B200000
B360000
D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L
Standard Fluids
Viscosity (cP)
5,000
12,500
30,000
60,000
Table - F1
Viscosity (cP) 25
Table - F2
High Temperature Standard Fluids
Three Viscosity/Temperatures**
HT-30,000
HT-60,000
HT-100,000
C
0
29
200
600
1,060
2,000
10,200
21,000
73,000
200,000
360,000
61