Qc Limits - Ametek BROOKFIELD DVPlus Operating Instructions Manual

Digital viscometer
Table of Contents

Advertisement

End conditions are defined in the table below:
Parameter
Time
Torque
Temperature
Viscosity
# of Revs
None
If "None" is selected, the DVPlus will run the test until stopped by the operator. The End Condition "Time" cannot be
set for more than 48 hours.
TIP: Be sure to select a time value that provides sufficient time to allow for the equilibrium of the torque sensor.
A short time value may result in erroneous data if the torque sensor has not achieved a stable deflection for the
measurement condition. AMETEK Brookfield recommends a minimum time value of 20 seconds. However, this
can be reduced based on user judgement by observing when the torque stabilizes.
TIP: Slow speeds require a longer time for the equilibrium of the measurement sensor. When using speeds less
than 5 RPM, consider a time value of 60 seconds or longer.

2.11 QC Limits

The DVPlus Viscometer has a QC Limits feature that allows the operator to select an acceptable range for
measurement results. The range may be defined by Torque, Viscosity, Temperature, and Time. The QC Limits feature
can also be turned off by selecting None.
M21-400 REV.1.1
Figure 2-10.1
Type
Units (per Settings)
Equal to (=)
Equal to (=)
Greater Than (>)
Equal to (=)
Less Than (<)
Greater Than (>)
Equal to (=)
Greater Than (>)
Equal to (=)
n/a
D V P L U S D I G I TA L V I S C O M E T E R - O P E R AT I O N A L M A N U A L
Figure 2-10.2
Seconds
1 second to 48 hours
%
mNm
dyne•cm
°C
°F
K
-100 to 300°C
cP
Pa•s
mPa•s
P
n/a
n/a
Range
Tolerance
0.0-100%
Varies
Varies
Varies
Varies
1-9999
n/a
n/a
n/a
n/a
30