CT 9000 ® ADV Digital Injection System
FACTORY CALIBRATION
From the Factory Calibration Screen you can Enable and Disable Face-
plates. (Refer to Figure 5-1a).
NOTE: The injector will display an error message if an injection is attempted with
a disabled faceplate. If a faceplate is disabled all calibration data for that
faceplate is lost.
Place the unit in the Calibration Mode (Refer to the previous Section—
Placing Unit in Calibration Mode). Remove the Pressure Plate Assembly
from the powerhead.
o
To activate the Factory Calibration Screen the appropriate pass-
word must entered. The password is activated by pressing a key
seguence. From the top key count down to key 4 and proceed as
follows:
Press key 4, then key 5, then key 4, then key 6.
The Factory Calibration Screen will be displayed.
O)
Press the desired Faceplate key to Enable or Disable. Your selec-
tion will toggle under the 'Enable' column either YES or NO.
α
Press the [Exit] key when selection is completed.
Move position 1 of SW1 on the CPU board to the "on" position. Then
press the [Exit] key again to return to normal operation..