Mallinckrodt CT 9000 ADV Installation And Service Manual page 18

Digital injection system
Table of Contents

Advertisement

CT 9000? ADV
NOTE:
The syringe heater for use with with the 200 ml (front-load) and
125 ml (pre-fill) type syringes with pressure sleeves is P/N 800114.
The syringe heater for push and turn type faceplate is P/N 800146
FLOW RATE PARAMETERS
200 ml Syringes
0.1—9.9 ml/s
125 ml Syringe
0.1
一 9.9 ml/s
100 ml Syringe
0.1—9.9 ml/s
30-50 ml Syringe
0.1—7.0 ml/s
PRESSURE LIMIT
Pressure Units
Range
pounds per square inch (psi)
25, 50—300 (150*)
(in 50 psi increments)
bar
2, 3, 7, 10*, 14, 17, 21
kg/em?
2, 4, 7, 11*, 14, 18, 21
NOTE:
(*) Used to denote pressure values pertaining only to the 200 ml
pushrod syringe
INJECT DELAY
Inject Delay Parameters
0—255 sec
SCAN DELAY
Scan Delay Parameters
0—99 sec
STORED PROTOCOLS
12 protocols can be stored and recalled.