Mallinckrodt CT 9000 ADV Installation And Service Manual page 117

Digital injection system
Table of Contents

Advertisement

Verify the Powerhead's Enable light is illuminated.
©
Press the [Start] key. Verify that the "Inject Time", rotating wheel,
and "Volume Delivered" are tracking correctly and that the
Powerhead's Injecting light is illuminated.
©
When the "Achieved Values" screen is displayed, verify the Time,
Flow, and Volume results are within tolerances as follows: The
pressure should be less than 300 psi in each case.
200 ml Pressure Plate Assembly:
Time = 20 + 1 second
Flow = 9.9 + 0.2 ml/sec
Volume = 195 +2 ml
125 ml Pressure Plate Assembly:
Time = 12.5 + 1 second
Flow = 9.9 + 0.2 ml/sec
Volume = 123 + 2 ml
200 ml Pushrod Pressure Plate Assembly:
Time = 20 + 1 second
Flow = 9.9 + 0.2 ml/sec
Volume = 195 + 2 ml
100 ml Pushrod Pressure Plate Assembly:
Time = 9.5 + 1 second
Flow = 9.9 + 0.2 ml/sec
Volume = 195 +2 ml
30-50 ml Hand-Held Pressure Plate Assembly:
Time = 4.25 + 1 second
Flow = 7.0 + 0.2 ml/sec
Volume = 30 +2 ml
800961-C
Nov. 2000
6-7