800961-C Nov. 2000
区 JALLINCKRODT
ENVIRONMENTAL
CAUTION!
The console and electronics cabinet are not spill-proof. Fluid spilled in
these components can cause the unit to malfunction, resulting in patient or
operator injury. If fluid is spilled on either component, remove from
operation and contact your authorized service personnel.
DANGER!
Possible explosion hazard if used in the presence of flammable anesthet-
ics. The unit is not designed for use in explosive environments.
WARNING!
The powerhead may only be operated in an area that is located beyond the
20 gauss limit. Operating the unit within magnetic fields that are higher
than this limit may cause the unit to malfunction, resulting in operator or
patient injury.
SYRINGES
Syringe Sizes
200 ml (front-load)
200 ml (pushrod)
125 ml (pre-filled)
100 ml (pushrod)
50 ml (hand-held)
30 ml (hand-held)
FILL RATE
Full forward or reverse time: 13-100 sec. Accelerates from zero
to maximum within approximately 3 seconds after pressing the
[Forward] or [Reverse] key in conjunction with one of the [Fast]
keys.
' SYRINGE HEATER
98° + 6° F (37° + 3° C) nominal. Maintains the temperature of
pre-heated contrast. Overtemp condition will occur at 107.6° F
(42° C).