CT 9000 ® ADV Digital Injection System
FLOW RATE CHECK
NOTE:
For the flow rate checks below, install the pressure kit used during the
=
calibration procedures in Chapter 5. Turn the valve 1/2 turn open.
©
Enter an Injection with the following parameters, for available
faceplates.
200 ml Pressure Plate Assembly:
Flow Rate:
9.9 ml/sec
Volume:
195 ml
Pressure Limit: 300 psi
125 ml Pressure Plate Assembly:
Flow Rate:
9.9 ml/sec
Volume:
123 ml
Pressure Limit: 300 psi
200 ml Pushrod Pressure Plate Assembly:
Flow Rate:
9.9 m/sec
Volume:
195 ml
Pressure Limit: 150 psi
100 ml Pushrod Pressure Plate Assembly:
Flow Rate:
9.9 ml/sec
Volume:
95 ml
Pressure Limit: 300 psi
30-50 ml Hand-Held Pressure Plate Assembly:
Flow Rate:
7.0 ml/sec
Volume:
30 ml
Pressure Limit: 300 psi
Enable the injection without saving.
人 の
Verify that the display will ask if air has been evacuated.
©
Press the [Yes] key. Expected Duration shall be between 4 and 20
seconds, depending on the type of pressure plate used.
200 ml Pressure Plate Assembly:
Expected Duration:
20 sec
125 ml Pressure Plate Assembly:
Expected Duration:
12 sec
200 ml Pushrod Pressure Plate Assembly:
Expected Duration:
20 sec
100 ml Pushrod Pressure Plate Assembly:
Expected Duration:
10 sec
30-50 ml Hand-Held Pressure Plate Assembly:
Expected Duration:
4-7 sec