Mallinckrodt CT 9000 ADV Installation And Service Manual page 83

Digital injection system
Table of Contents

Advertisement

区 JALLINCKRODT
Overheating error.
To continue, press override. See manual for further instructions.
(See also Fault Message # 23)
Occurs when: A problem with the heater blanket has been de-
tected,
Operator action: Press [Override], power down the unit and discon-
nect the heater blanket. Discontinue use of the heater blanket until
serviced by a qualified Liebel-Flarsheim service technician. The CT
9000 is completely functional, however, the temperature of heated
contrast cannot be maintained.
ERROR CODES
E0
Occurs when: A pressure plate has not been installed and the
operator presses the [AutoFill] key.
Operator action: Install the appropriate pressure plate prior to
activating the AutoFill function.
E2
Occurs when: With a pressure plate installed, the operator presses
the [AutoFill] key, however the ram has not been moved far enough
for the AutoFill function to be activated.
Operator action: Continue to move the ram to until it is fully ex-
tended.
FAULT MESSAGES
The dual microprocessor control circuits in the CT 9000 continu-
ously monitor all important functions from the moment that the unit
is turned on. These checks, transparent to the operator, occur
many times each second, monitoring the safe performance of the
injector. If a fault is detected, the unit will lock into the Fault Mode
and then display a Fault Message. Refer to Figures 4-1 and 4-2.
In all cases, the appearance of a Fault Message requires that the
operator contact a qualified service technician to correct the prob-
lem. A listing of these Fault Messages, along with its corrective
action, follows.
800961-C Nov. 2000
4-3

Advertisement

Table of Contents
loading

Table of Contents