Installing the
aerosol cover
Connecting the
power supply cable
Thermo Fisher Scientific
2. Additionally, add a commercially available anti-foaming agent, such
as silicone oil, to the empty waste container to reduce foaming. Use
concentrations of anti-foaming agents as recommended by the
manufacturers.
3. If foaming continues to be a problem, we recommend switching to a
larger waste container and additionally increasing the concentration
of anti-foaming agent in the waste container.
4. To help facilitate breakdown of the foam in the waste container,
carefully swirl the waste container from time to time to improve
mixing between foam layer and anti-foaming agent.
During operation, the see-through aerosol cover can be present
(Figure 2–2) or absent. The aerosol cover primarily protects the user
against biohazardous aerosols and also against environmental
contamination. The aerosol cover reduces the amount of airborne
pathogens making the air you breathe healthier and cleaner.
Figure 3–20 shows the aerosol cover being installed. The corners have
magnets for quick magnetic mounting of the cover.
Figure 3–20. Magnetic mounting of the aerosol cover
To connect the power supply cable:
Warning Never operate your instrument from a power outlet that has
no ground connection. Never use a power supply cable other than the
Thermo Scientific power supply cable designed for your region.
Thermo Scientific Wellwash Versa User Manual
Installation
Installation setups
31