8 After use
8.1 After each treatment session
64
8
After use
8.1
After each treatment session
NOTICE
Perform reprocessing immediately after treatment, after 1 hour at the
latest.
IMPORTANT
Cable inspection
Inspect the cable of the charger, the handpiece, the lip clip, and the file
clamp before and after each use. If you detect wear or tear of the
cables, contact the service center.
NOTICE
Use only Dentsply Sirona T1 Spray.
ü Wear suitable protective clothing.
1. Remove the file with tweezers.
2. Detach the contra-angle handpiece from the handpiece sleeve.
3. Detach the handpiece sleeve from the handpiece.
4. Detach the lip clip cable and file clamp from the apex locator
adapter.
5. Detach the lip clip from the lip clip cable.
6. Peform the "Reprocessing [→ 65]".
8.2
At the end of the work day
➢ Lubricate the contra-angle handpiece with spray [→ 68].
NOTICE
Do not leave a contra-angle handpiece on the straight handpiece
overnight to prevent oil from leaking into the electric motor. Never
lubricate the electric motor.
8.3
Once per work week
ü The motor has been conditioned.
➢ Lubricate the lock washer [→ 70].
Dentsply Sirona
Instructions for Use X-Smart
67 98 487 D3778
D3778.201.01.02.02 2023-05