2. Assemble the water tank by inserting the handle, the lid and the
04:00 PM
float including the float guide.
3. If you do not have an automatic water connection for your DAC
04:00 PM
Universal, fill demineralized/distilled water (water conductivity < 3.0
µS/cm) into the water tank (min. 1.0 liter, max. 2 liter).
4. Insert the water tank back into the DAC Universal.
04:00 PM
IMPORTANT
When inserting make sure that the guide hooks on the top rear side of
the water tank are inserted correctly into the housing.
5.1.21
Preparations after the Check & Clean process
Frequency: Weekly or every 50 cycles (whichever comes first).
ü The DAC Universal is switched on and is in the Check & Clean
process menu.
ü NitraClean cleaning tablets are available.
CAUTION
Risk of poisoning!
Only use the NitraClean cleaning tablets stated below
1. Go to the next Check & Clean option by pressing the " Next Check &
Clean step " function key.