3 System description
3.4 Electromagnetic compatibility
Phenomenon
Electrostatic discharge
RF electromagnetic fields
RF electromagnetic fields in the immedi-
ate vicinity of wireless communications
equipment
Rated power frequency magnetic fields
Magnetic fields in the vicinity
Phenomenon
Input AC power port
Surges
Line to line
Surges
Line to ground
Conducted disturbances induced by RF
fields
Voltage dips
Voltage interruptions
20
3.4.2

Interference immunity

Enclosure
Basic EMC standard
or test method
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
IEC 61000-4-39
Input AC power port
Basic EMC standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Instructions for Use X-Smart
Interference immunity test level for profes-
sional healthcare facilities
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
See table "Test specifications for the immunity
of enclosure ports to RF wireless communica-
tions equipment"
30 A/m
50 Hz or 60 Hz
65 A/m – 134.2 kHz (2.1 kHz PM)
7.5 A/m – 13.56 MHz (50 kHz PM)
Interference immunity test level for profes-
sional healthcare facilities
± 2 kV
100 kHz repetition frequency
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2 kV
3 V
0.15 MHz to 80 MHz
6 V in ISM bands between 0.15 MHz and
80 MHz
80% AM at 1 kHz
0% UT; 1/2 cycle
at 0°, 45°, 90°, 135°, 180°, 225°, 270° and
315°
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0°
0% UT; 250/300 cycles
D3778.201.01.02.02    2023-05