9 Electromagnetic compatibility

9.2 Interference immunity

Interference
IEC 60601-1-2 test level
immunity tests
Electrostatic
± 6 KV contact
discharge
discharge (ESD)
± 8 KV air discharge
according to IEC
61000-4-2
Electrical fast
± 1  k V for input and
output lines
transient/burst
± 2 kV for power supply
lines
according to
IEC 61000-4-4
Surge voltages
± 1 kV differential mode
according to IEC
± 2 kV common mode
61000-4-5
voltage
Voltage dips, short
<5% U
interruptions and
(>95% dip of U
variations of the
40% U
power supply
(60% dip of U
according to IEC
70 % U
61000-4-11
(30% dip of U
<5% U
(>95% dip of U
Magnetic field of
3 A/m
power frequencies
(50/60 Hz) according
to IEC 61000-4-8
Remarks: U
is the AC supply voltage prior to application of the test level.
T
74
Interference immunity
9.2
The UNIT is intended for operation in the electromagnetic environment
specified below.
The customer or user of the UNIT should make sure that it is used in
such an environment.
Compliance level
± 6 KV contact
discharge
± 8 KV air discharge
± 1  k V for input and
output lines
± 2 kV for power
supply lines
± 1 kV differential
mode
± 2 kV common mode
voltage
for ½ period
<5% U
T
)
(>95% dip of U
T
for 5 periods
40% U
T
)
(60% dip of U
T
for 25 periods
70 % U
T
)
(30% dip of U
T
for 5sec.
<5% U
T
(>95% dip of U
T
3 A/m
Operating Instructions SiroLaser Blue
Electromagnetic environment –
guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
The quality of the line power supply
should be that of a typical commer-
cial or hospital environment.
The quality of the line power supply
should be that of a typical commer-
cial or hospital environment.
for ½ period
The quality of the line power supply
T
)
should be that of a typical commercial
T
or hospital environment.
for 5 periods
T
)
If the user of the UNIT requires it to
T
continue functioning following interrup-
for 25 periods
T
tions of the power supply, it is recom-
)
T
mended to have the UNIT powered by
for 5sec.
T
an uninterruptible power supply or a
T
battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
Portable and mobile radio equip-
ment must not be used within the
recommended working clearance
from the UNIT and its cables, which
is calculated based on the equation
suitable for the relevant transmission
frequency.
Recommended working clearance:
Dentsply Sirona
66 56 842 D3648
D3648.201.03.02.02    10.2018