Interference immu-
IEC 60601-1-2 Test level
nity tests
Electrostatic dis-
± 4 kV contact
charge (ESD) ac-
± 8 kV air
cording to IEC
61000-4-2
Electrical fast tran-
± 0.5 kV for input and
sient/burst accord-
output lines
ing to
± 1 kV for power supply
IEC 61000-4-4
lines
Surge voltages
± 0.5  k V differential mode
according to IEC
voltage
61000-4-5
± 1 kV common mode
voltage
Voltage dips,
0% U
short interruptions
(100% dip of U
and variations of
the power supply
according to
IEC 61000-4-11
Conducted RF in-
1 V
terference
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF inter-
3  V /m
ference
80MHz to 1 GHz
IEC 61000-4-3
3  V /m
1.4  G Hz to 2.0 GHz
1  V /m 1.4 GHz to 2.7  G Hz
65 71 165 D3639
D3639.201.01.08.02    01.2020
3.5.2

Interference immunity

The UNIT is intended for operation in the electromagnetic environment
specified below.
The customer or user of the UNIT should make sure that it is used in
such an environment.
Compliance level
± 4 kV contact
± 8 kV air
± 0.5 kV for input and
output lines
± 1 kV for power supply
lines
± 0.5  k V differential
mode voltage
± 1 kV common mode
voltage
for ½ period
0% U
T
)
(100% dip of U
T
1 V
eff
3 V/m
3 V/m
1 V/m
Remark 1
The higher frequency range applies at 80 MHz and 800 MHz.
Electromagnetic environment -
guidelines
Floors should be wood, concrete, or
ceramic tile. If the floor is covered
with synthetic material, the relative
humidity must be at least 30%.
The quality of the line power supply
should be that of a typical commer-
cial or hospital environment.
The quality of the line power supply
should be that of a typical commer-
cial or hospital environment.
for ½ period
The quality of the line power supply
T
)
should be that of a typical commer-
T
cial or hospital environment.
d= [1.2] √P
eff
d= [1.2] √P
at 80  M Hz to 1  G Hz
d= [2.3] √P
at 1.4  G Hz to 2.7  G Hz
with P as the power rating of the
transmitter in watts (W) according to
the transmitter manufacturer's speci-
fications and d as recommended
safety distance in meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey
than the compliance level
frequency range.
Interference is possible in the vicinity
of equipment bearing the following
graphic symbol.
3 Product description
3.5 Electromagnetic compatibility
1
should be less
2
in each
27