Custom Setting - Horiba Scientific LAQUA F-73A-S Instruction Manual

Ph/ion meter
Table of Contents

Advertisement

Each HOLD condition is described below.
Stability condition
Auto hold
Measuring
Mode
target
pH
EXACT
ION, mV, ORP
pH
NORMAL
ION, mV, ORP
pH
BRIEF
ION, mV, ORP
TIME
Common
pH
CUSTOMI
ION
ZE
mV
ORP
Manual hold

2.3 Custom Setting

18
In the AUTO HOLD mode, the instrument judges potential stability
automatically to set the measurement values.
Temperature
Time (s)
(C)
10
2.0
10
2.0
10
2.0
-
-
Arbitrary
setting
2.0
2 s to 60 s
【10 s】
Determine an end point manually.
(Tap
to hold it.)
We will explain the procedures of CUSTOMIZE
setting taking the AUTO HOLD item as an exam-
ple.
1.
Select the CUSTOMIZE of the Hold type to set
the stability condition time and the stability
condition value.
2.
Use the numeric-key screen to enter
measurement variations as HOLD criteria for
each measurement item.
Tap
Function
Content
Criteria
0.005 pH (Equivalent to 0.3 mV)
0.3 mV
0.015 pH (Equivalent to 1.0 mV) Default setting of
1.0 mV
0.050 pH (Equivalent to 3.0 mV)
3.0 mV
Arbitrarily set at 2 s to 999 s.
Arbitrarily set at 0.001 pH to
0.100 pH.
Arbitrarily set at 0.1 mV to 60 mV.
Arbitrarily set at 0.1 mV to 60 mV.
Arbitrarily set at 0.1 mV to 60 mV.
to return to the previous screen.
【Default】
auto hold
【10 s】
【0.005 pH】
【0.3 mV】
【0.3 mV】
【0.3 mV】