5.2.1 Solution Temperature Entry
in MTC (Manual Temperature Setting)

5.3 Alarm Setting

When the measurement values exceed the set upper or lower limit, the instrument detects
it to display the notice on the screen or to output the signal from the external output
terminal.
If the measurement values exceed the alarm range, the color of the pertinent channel
"CH" is changes on the MEAS screen.
Set the upper limit alarm to ON for the upper limit control of measurement value.
Set the lower limit alarm to ON for the lower limit control of measurement value.
Upper limit value
64
1.
Display the Temperature item when select MTC.
Tap
on the right of the Temperature item.
2.
Enter the solution temperature on the numerical-
key screen.
3.
Tap
.
The setting applies.
To cancel the settings, tap
Lower limit value
.