Quick Manual
pH Meter F-72,73,74
pH / ION / COND METER F - 74
CAL
MEAS
CODE : GZ0000272694B
January, 2017
© 2017 HORIBA Advanced Techno Co., Ltd.
① Prepare pH standard solutions
② Connect the electrode to the pH meter
③ Press the power button of the pH meter
④ Remove the protective cap from the electrode,
DATA
Remove the protective cap.

Preparation

pH/ ION / COND METER F-74
Electrode terminal
CAL
MEAS
DATA
and open the internal solution ller port
4
4
9
9
pH
pH
pH
pH
Temperature sensor terminal
☆ Press the power button for 2 seconds.
※ Wait for 10 seconds after connecting
  the AC adapter before pressing
  the power button.
Open the internal solution ller port.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LAQUA F-72 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Horiba Scientific LAQUA F-72

  • Page 1 Preparation ① Prepare pH standard solutions Quick Manual pH Meter F-72,73,74 ② Connect the electrode to the pH meter pH/ ION / COND METER F-74 Electrode terminal Temperature sensor terminal pH / ION / COND METER F - 74 ③ Press the power button of the pH meter ☆...
  • Page 2 Calibration Measurement When the measurement sample is known to be acid, perform two-point calibration with pH7 and pH4. When it is known to be alkaline, perform two-point calibration with pH7 and pH9. When the sample pH is not known, perform three-point calibration with pH4, pH7 and pH9. ①...

This manual is also suitable for:

Laqua f-73Laqua f-74

Table of Contents