Checking Before Use - Horiba Scientific LAQUA F-73A-S Instruction Manual

Ph/ion meter
Table of Contents

Advertisement

・For calibration using multiple standard solutions, continuously perform from one solution
to another. If returned to the MEAS screen once, you have to go back to the 1st point
calibration.
・When an error occurs in calibration, the screen displays the error.
Remove the cause and start the calibration from the beginning.

4.3 Checking Before Use

When the preliminary check function is ON, you can check repeatability with the standard
solution of the 1st point calibration. Use this function to check measurement accuracy by
checking repeatability.
An operation guide is displayed after the 2nd point calibration is completed. Check the
repeatability according to the operation guide.
After the set point calibrations are completed, repeatability is automatically checked using
the deviation between the calibrated and measured values at the 1st point.