Custom Setting - Horiba Scientific DS-72G Instruction Manual

Cond meter
Table of Contents

Advertisement

Each HOLD condition is described below.
Stability condition
Auto hold
Measuring
Mode
target
COND, Resist
EXACT
SAL
TDS
COND, Resist
NORMAL
SAL
TDS
COND, Resist
BRIEF
SAL
TDS
TIME
Common
COND
CUSTOMI
SAL
ZE
Resist
TDS
Manual hold

2.3 Custom Setting

In the AUTO HOLD mode, the instrument judges potential stability
automatically to set the measurement values.
Temperature
Time (s)
(C)
10
2.0
10
2.0
10
2.0
-
-
Arbitrary
setting
2.0
2 s to 60 s
【10 s】
Determine an end point manually.
(Tap
to hold it.)
We will explain the procedures of CUSTOMIZE
setting taking the AUTO HOLD item as an exam-
ple.
1.
2.
Function
Content
Criteria
Minimum display digit: 1 digit
0.30 ppt (0.03%)
10 mg/L
Minimum display digit: 3 digits
1.00 ppt (0.10%)
30 mg/L
Minimum display digit: 5 digits
3.00 ppt (0.30%)
100 mg/L
Arbitrarily set at 2 s to 999 s.
Arbitrarily set at 0.001 mS/cm to
0.100 mS/cm (0.1 mS/m to 10.00
mS/m).
Arbitrarily set at 0.10 PPT to
10.00 PPT (0.01% to 1.00%).
Setting value of COND is reflected.
Arbitrarily set at 0.1 mg/L to 100
mg/L.
Select the CUSTOMIZE of the Hold type to set
the stability condition time and the stability
condition value.
Use the numeric-key screen to enter
measurement variations as HOLD criteria for
each measurement item.
Tap
to return to the previous screen.
【Default】
Default setting of
auto hold
【10 s】
【0.001 mS/cm
(0.1 mS/m)】
【0.3 PPT
(0.03%)】
【0.1 mg/L】