Orp Measurement Setting; Temperature Setting; Solution Temperature Entry In Mtc (Manual Temperature Setting) - Horiba Scientific LAQUA F-73A-S Instruction Manual

Ph/ion meter
Table of Contents

Advertisement

6.2 ORP Measurement Setting

This section describes the procedures to set the conditions of ORP measurement.
The setting procedures for each item are explained below.

6.3 Temperature Setting

If the temperature terminals of the instruction and electrode are not connected,
temperature setting is performed in MTC even when ATC is set.
6.3.1 Solution Temperature Entry
in MTC (Manual Temperature Setting)
70
1.
Tap the channel setting and the measurement
item in the MEAS screen to set "CH1" and
"ORP".
2.
Tap
and tap "CH1 MEAS SET".
3.
ORP measurement setting items are displayed.
When setting item increased, you will see the
remaining items by dragging.
4.
Select items and set the conditions.
There are two types of temperature setting for
ORP measurement; Automatic Temperature
setting (ATC) and Manual Temperature setting
(MTC).
In ATC, the instrument detects the solution
temperature with the connected temperature
sensor, and displays it on the screen.
In MTC, measure the solution temperature and
enter the value in advance. The instrument
displays the entered temperature.
1.
Display the Temperature item when select MTC.
Tap
on the right of the Temperature item.
2.
Enter the solution temperature on the numerical-
key screen.
3.
Tap
.
The setting applies.
To cancel the settings, tap
.