Chapter 2 Before Measurement (Meter SET)

This chapter explains the procedures of the instrument condition setting, which should be
performed before measurement.

2.1 Meter SET Screen

2.2 Auto Hold Setting

1.
Tap
and tap Meter SET.
Meter SET items are displayed.
You will see the remaining items by dragging.
2.
Select items and set the conditions.
The setting procedures for each item are
explained below.
In the AUTO HOLD mode, the instrument judges
potential stability automatically to the
measurement values. This instrument allows
you to select one among the 6 type criteria of
potential stability.
1.
Change the auto hold settings, tap
right of the AUTO HOLD item.
2.
Tap
on the right of the HOLD TYPE item.
3.
Select the measurement stability condition of the
6 types (EXACT, NORMAL, BRIEF, TIME,
CUSTOMIZE, Manual) in the AUTO HOLD
selection screen.
To cancel the operation, tap
previous screen.
on the
to return to the