Measurement; Sample Setting - Horiba Scientific LAQUAtwin B-74 Series Manual

Table of Contents

Advertisement

6

Measurement

6.1

Sample setting

The following 4-type sampling setting methods are available.
 Drop: for a small-amount sample
 Immersion: for a large-amount sample
 Scooping: for sampling a part of a sample
 Sampling sheet: for a sample containing tiny particles
Note
Although this product is waterproof, avoid immersing it completely.
If the product is accidentally dropped into water, take it out of water
and wipe off the moisture on it.
 Drop
1.
Open the light shield cover and put some drops of sample on the
flat sensor to cover the entire flat sensor.
2.
Close the light shield cover.
Light shield cover
Tip
Try to use the provided sampling sheet B
for a minute sample.
Using this sheet, the entire flat sensor can
be covered with only 50 L to 100 L
sample.
 Note that a reaction between a
sample and the sampling sheet B
may affect the measured value.
 Handle the sampling sheet B with tweezers to minimize possible
contamination.
 Make sure to close the light shield cover during measurement to
minimize possible sample evaporation.
13