Chapter 4 ION Measurement

4.1 ION Calibration Setting

This section describes the procedures to set the conditions of ION calibration.

4.1.1 Calibration Points Setting

1.
Tap the channel setting and the measurement
item in the MEAS screen to set "CH1" and "ION".
2.
Press the CAL key to display the ION CAL
screen.
3.
Tap
and tap "CH1 CAL SET".
Ion calibration setting items are displayed.
4.
Select items and set the CONDITIONS.
The setting procedures for each item are
explained below.
Up to 5 points calibration is possible.
Set the number of the calibration points.
1.
Tap
on the right of the Calibration Points
item.
2.
Tap
or
in the Calibration points setting
screen to set the calibration points.
3.
Tap
.
The setting applies.
To cancel the settings, tap
.