10.3.2 Measurement Condition Default Settings
(Can Be Set per Operator)
Alarm
condition
Resolution
pH
Temperature
measurement
setting
condition
Temperature
conversion
Electrode
data
Standard solution
Calibration points
Calibration interval setting
pH calibration
condition
Calibration interval
Preliminary check
Periodical check
Item
Upper limit
value setting
Lower limit
value setting
Upper limit
value
Lower limit
value
Temperature
setting
Temperature
input value in
selecting "MTC"
Temperature
conversion
function
Temperature
conversion
coefficient
Standard
temperature
Model
lot No.
Selection item/
Setting range
Enable, Disable
Enable, Disable
0.000 pH to 14.000 pH
0.000 pH to 14.000 pH
0.01, 0.001, AUTO
ATC (Automatic temperature
compensation),
MTC (Manual temperature
compensation)
0.0C to 100.0C
Enable, Disable
±0.100 pH/C
0.0C to 100.0C
NIST, NIST2, USA, China,
CUSTOM
1 point to 5 points
Enable, Disable
1 day to 999 days
Enable, Disable
JIS, Pharmacopeia,
X-51 (Digital simulator)
Default
values
Disable
Disable
14.000 pH
0.000 pH
0.001
ATC
25.0C
Disable
0.000 pH/C
0.0C
None
None
USA
2 points
Disable
3 days
Disable
X-51
97