4.1.2 Checking Before Use

52
If you set the Checking Before Use setting to
ON, it allows you to check repeatability after
calibration by comparing the measured value of
first point calibration standard solution with the
calibration result.
After the set point calibrations are completed,
repeatability is automatically checked using the
deviation between the calibrated and measured
values at first point calibration.
For example, the sequence for the 2 points cali-
bration is as follows.
Calibration of 1st point --> Calibration of 2nd
point --> Repeatability check with the calibrated
value by 1st point
1.
Tap
on the right of the Checking Before Use
item to set the preliminary checking.
2.
Tap
.