8.2 ION Periodic Inspection Mode Setting
In the ION periodical check mode, the instrument operations are checked by inputting the
output of the optional checker X-51 into the instrument.
Refer also to the instruction manual of the checker X-51 before the operation.
Set the measurement unit for ion to mol/L before starting the check operation.
After the setting change, the settings apply.
Follow the guidance to calibrate the instrument by simulated input of 0 mV and 120 mV in
the order below and check the following operation.
Simulated calibration check
0 mV
1.00 mol/L
120 mV
10.0 mmol/L
Simulated linearity check
100 mol/L
240 mV
120 mV
10.0 mmol/L
0 mV
1.00 mol/L
120 mV 100 mol/L
Simulated temperature check
0.0C
30.0C
60.0C
100.0C
When all the check is completed, the result is outputted automatically.
89