3
100 µs
4
200 µs
5
500 µs
6
1 ms
7
2 ms
8
5 ms
9
10 ms
10
20 ms
11
50 ms
12
100 ms
13
200 ms
14
500 ms
15
1 s
N/A means that n is an illegal value under these conditions.
The TC. command is only used for reading the time constant, and reports the current
setting in seconds. Hence if a TC 12 command were sent and the noise measurement
mode was turned off, TC would report 12 and TC. would report 1.0E-01, i.e. 0.1 s or
100 ms.
SYNC [n]
Synchronous time constant control
n
Effect
0
Synchronous time constant disabled
1
Synchronous time constant enabled
SLOPE [n]
Output low-pass filter slope (roll-off) control
When the fast analog output mode and Noise Measurement mode are turned off
(FASTMODE 0 and NOISEMODE 0), the selection is according to the following
table:
n
Slope
0
6 dB/octave
1
12 dB/octave
2
18 dB/octave
3
24 dB/octave
When:
a) The fast output filter mode is turned on, or
b) Noise measurement mode is turned ON (NOISEMODE 1)
the selection is according to the following table:
n
Slope
0
6 dB/octave
1
12 dB/octave

6.7.04 Signal Channel Output Amplifiers

XOF [n
[n
]]
X channel output offset control
1
2
n
Selection
1
0
Disables offset
1
Enables offset facility
The range of n
is ±30000 corresponding to ±300% full-scale.
2
YOF [n
[n
]]
Y channel output offset control
1
2
Chapter 6, COMPUTER OPERATION
N/A
19
N/A
20
500 µs
21
1 ms
22
2 ms
23
5 ms
24
10 ms
25
N/A
26
N/A
27
N/A
28
N/A
29
N/A
30
N/A
20 s
N/A
50 s
N/A
100 s
N/A
200 s
N/A
500 s
N/A
1 ks
N/A
2 ks
N/A
5 ks
N/A
10 ks
N/A
20 ks
N/A
50 ks
N/A
100 ks
N/A
6-15