Chapter 5, FRONT PANEL OPERATION

5.5.02 Dual Reference Main Display

5-62
Once dual reference has been activated, the Main Display - Dual reference Mode is
displayed, as shown in figure 5-49.
Figure 5-49, Main Display - Dual Reference Mode
In dual reference mode, controls and displays relating to the reference channel 1
signal carry the suffix "1" and those for reference channel 2, the suffix "2". The
Reference Channel Menu 1 is used to select which reference source is used for each
reference channel. Hence, for example, if the reference channel 2 is set to Internal
mode and reference channel 1 is set to External Front Panel, an external reference
frequency of 1 kHz is applied to the main console's front panel connector and the
instrument's internal oscillator is set to 15 kHz, the unit will measure signals at both
1 kHz and 15 kHz. The Sensitivity 1, Time Constant 1 and Ref Phase 1 controls will
affect the signal being measured at 15 kHz, while the Sensitivity 2, Time Constant 2
and Ref Phase 2 controls will affect the signal being measured at 1 kHz.
The range of controls that can be placed on the left-hand side of the Main Display
using the SELECT CONTROL key is therefore extended to include those relating to
both reference signals, as follows:-
Figure 5-48, Dual reference Setup Menu