n
Input configuration
0
Both inputs grounded (test mode)
1
A input only (remote connection unit)
2
-B input only (remote connection unit)
3
A-B differential mode (remote connection unit)
4
AUX input (main console)
Note that the IMODE command takes precedence over the VMODE command.
VRLOCK [n]
Virtual reference mode lock
The Seek option of the frequency sweep mode must be used before issuing this
command, for which the value of n has the following significance:
n
Mode
0
Disables virtual reference mode
1
Enters virtual reference mode by enabling tracking of the signal frequency
X[.]
X channel output
In fixed point mode causes the lock-in amplifier to respond with the X demodulator
output in the range ±30000, full-scale being ±10000.
In floating point mode causes the lock-in amplifier to respond with the X
demodulator output in volts or amps.
XOF [n
[n
]]
X channel output offset control
1
2
n
Selection
1
0
Disables offset
1
Enables offset facility
The range of n
is ±30000 corresponding to ±300% full-scale.
2
XY[.]
X, Y channel outputs
Equivalent to the compound command X[.];Y[.]
Y[.]
Y channel output
In fixed point mode causes the lock-in amplifier to respond with the Y demodulator
output in the range ±30000, full-scale being ±10000.
In floating point mode causes the lock-in amplifier to respond with the Y
demodulator output in volts or amps.
YOF [n
[n
]]
Y channel output offset control
1
2
n
Selection
1
0
Disables offset facility
1
Enables offset facility
The range of n
is ±30000 corresponding to ±300% full-scale
2
Appendix F, ALPHABETICAL LISTING OF COMMANDS