The Curve Trigger menu is used to control how trigger inputs affect the curve buffer,
and the position and polarity of the trigger output.
ACQUISITION TRIGGER
The acquisition trigger is a "per curve" (start) or "per point" (sample) trigger
detecting the signal at the TRIG IN connector on the rear panel of the main console,
as follows:
INTERNAL
The curve buffer acquisition is not affected by input triggers
EXT
PER CURVE
Acquisition starts on detection of a rising edge on the TTL trigger signal. It
continues until the stop trigger condition is satisfied; each data point is taken at
the buffer time per point setting.
EXT
PER POINT
Acquisition starts on detection of a rising edge on the TTL trigger signal, with
each data point requiring a new trigger. It continues until the stop trigger
condition is satisfied; the buffer time per point setting has no effect.
EXT
PER CURVE
Acquisition starts on detection of a falling edge on the TTL trigger signal. It
continues until the stop trigger condition is satisfied; each data point is taken at
the buffer time per point setting.
EXT
PER POINT
Acquisition starts on detection of a falling edge on the TTL trigger signal, with
each data point requiring a new trigger. It continues until the stop trigger
condition is satisfied; the buffer time per point setting has no effect.
STOP TRIGGER
This control sets the condition used to terminate acquisition into the curve buffer, as
follows:
INTERNAL
The curve buffer acquisition continues until stopped manually from the front
panel or via an HC (halt curve) computer command
Chapter 5, FRONT PANEL OPERATION
Figure 5-33, Curve Trigger Menu
5-43