Usb Status Menu - Ametek 7124 Instruction Manual

Precision lock-in amplifier
Table of Contents

Advertisement

Chapter 5, FRONT PANEL OPERATION

5.3.19 USB Status Menu

5-34
When the Configuration menu is displayed, pressing a key adjacent to the USB
status item accesses the USB Status menu, which is shown in figure 5-25.
This menu has five indicators reporting the USB computer interface status and a key
for accessing the Communications Monitor display, as follows:-
Manufact. Descriptor
This is the manufacturer text string as reported in the USB descriptor.
Product Descriptor
This is the product description text string as reported in the USB descriptor.
Ser. Num. Descriptor
This is the instrument serial number text string as reported in the USB descriptor.
STATUS
This indicates whether the instrument is connected to the USB and has been
correctly enumerated by the controlling computer.
CONNECTION SPEED
If the USB status is connected, then this indicator shows the actual connection speed.
OUTPUT TERMINATOR
This has two settings, as follows:
NULL
In this setting the ASCII character responses that the instrument sends returns via
the USB interface are terminated in a null character
STATUS
In this setting the ASCII character responses that the instrument sends returns via
the USB interface are terminated with the Status and Overload bytes.
Pressing the Communications Monitor key accesses the Communications Monitor
display, while pressing the Previous Menu key returns control to the Configuration
Menu.
Figure 5-25, USB Status Menu