Chapter 6, COMPUTER OPERATION
AS
ASM
ACGAIN [n]
AUTOMATIC [n] AC Gain automatic control
LF [n

6.7.02 Reference Channel

REFMODE [n]
6-12
Floating point mode can only be used for reading the sensitivity, which is reported in
volts or amps. For example, if IMODE = 0 and the sensitivity is 1 mV the command
SEN would report 18 and the command SEN. would report +1.0E-03. If IMODE was
changed to 1, SEN would still report 18 but SEN. would report +1.0E-09
Perform an Auto-Sensitivity operation
The instrument adjusts its full-scale sensitivity so that the magnitude output lies
between 30% and 90% of full-scale.
Perform an Auto-Measure operation
The instrument adjusts its full-scale sensitivity so that the magnitude output lies
between 30% and 90% of full-scale, and then performs an auto-phase operation to
maximize the X channel output and minimize the Y channel output.
AC Gain control
Sets the gain of the signal channel amplifier according to the following table:-
n
AC Gain
0
0 dB
1
6 dB
2
12 dB
3
18 dB
4
24 dB
5
30 dB
6
36 dB
7
42 dB
n
Status
0
AC Gain is under manual control, either using the front panel or the ACGAIN
command
1
Automatic AC Gain control is activated, with the gain being adjusted according
to the full-scale sensitivity setting
n
]
Signal channel line frequency rejection filter control
1
2
The LF command sets the mode and frequency of the line frequency rejection
(notch) filter according to the following tables:
n
Selection
1
0
Off
1
Enable 50 or 60 Hz notch filter
2
Enable 100 or 120 Hz notch filter
3
Enable both filters
n
Notch Filter Center Frequencies
2
0
60 Hz (and/or 120 Hz)
1
50 Hz (and/or 100 Hz)
Reference mode selector
The value of n sets the reference mode of the instrument according to the following
table:
n
AC Gain
8
48 dB
9
54 dB
10
60 dB
11
66 dB
12
72 dB
13
78 dB
14
84 dB
15
90 dB