Ametek 7124 Instruction Manual page 99

Precision lock-in amplifier
Table of Contents

Advertisement

Chapter 5, FRONT PANEL OPERATION
Figure 5-34, Curve Select Menu - Standard Mode
Figure 5-35, Curve Select Menu - Fast Mode
For the standard mode, this menu is used to select the instrument outputs that will be
stored when data acquisition to the curve buffer is initiated. The possible data types
(outputs) that can be stored to the curve buffer are shown on the upper part of the
screen. Controls allow between one and all of these data types to be selected for
storage, with those that are selected being indicated by being shown in highlighted
text.
MOVE POINTER
This control allows the selection box to be moved to any one of the possible data
types. The
,
,
and
cursor-movement keys can be also be used to do this.
ENTER SELECTION / CLEAR SELECTION
If the data type enclosed by a selection box is not selected, then pressing this key
causes it to be selected, as indicated by its being displayed in bright text. If it is
already selected, then pressing the key deselects it.
As shown in figure 5-35, when the buffer is used in fast mode, five output values are
stored and there is no provision for changing these.
Pressing the Previous Menu key returns control to the Curve Buffer menu.
5-45

Advertisement

Table of Contents
loading

Table of Contents