Error Codes
Error codes are shown in the main display for approx. 2 seconds, after which the program automati-
cally returns to the weighing mode.
Display
No segments appear on
the display
h
l
e 01
e 02
E 09
e 10
e 11
e 22
e 30
Max. weighing range is less
than indicated under
"Specifications"
The weight readout
is obviously wrong
If any other errors occur, contact your local Sartorius Service Center!
22
Cause
No AC power is available
The AC adapter is not
plugged in
Battery is dead
The load exceeds the
balance/scale capacity
The load plate is not in place
Something is touching
the load plate
Display capacity exceeded:
Value to be output cannot be
shown on the display
Calibration parameter
not met; e.g.:
– balance/scale not zeroed
– balance/scale is loaded
If the gross weight ² zero,
no tare weight is available
The t key is blocked
when there is data in
the second tare memory
(net-total) – only 1 tare function
can be used at a time
Value input is not allowed for
second tare memory
Weight is too light or there is
no sample on the balance/scale
Interface port for printer
output is blocked
The balance/scale was
turned on without the weighing
pan in place
The balance/scale has not
been calibrated/adjusted
The balance/scale was not
tared before weighing
The balance/scale is not level
Solution
Check the AC power supply
Plug in the AC adapter
Replace the battery
Recharge the battery using
an external recharging device
Unload the balance/scale
Place the load plate on the
balance/scale
Move the object that is
touching the load plate
Decrease the weight on
the scale
Unload the balance/scale
Press t to tare the
balance/scale
Calibrate only when zero
is displayed
Tare the balance/scale
Press c to clear the
tare memory and release the
tare key
Press t
Increase the reference quantity
Contact your local Sartorius
Service Center
Place the weighing pan on
on the balance/scale
and press e to turn on the
balance/scale
Calibrate/adjust the
balance/scale
Tare before weighing
Level the balance/scale
Need help?
Do you have a question about the Basiclite and is the answer not in the manual?