Calibration/
Adjustment*
Always calibrate/adjust
the balance/scale after
setting it up in a new
location.
Available Features
Calibration/adjustment
can only be performed
when:
* "Calibration" technically means to determine the difference between the
balance/scale readout and the actual weight on the pan to determine the
accuracy. Adjustment means to bring a balance/scale into the state of
accuracy required for its use. Therefore, "calibration," as used in this manual,
actually means "adjustment."
10
– There is no load on
the balance/scale
– The balance/scale
is tared
– The internal signal
is stable
If these conditions are not
met, an error message is
displayed. The weight
required for calibration/
adjustment is displayed.
Calibration/Adjustment
for Verified Balances/
Scales of Accuracy
Class K and m
– The external calibra-
tion function is
blocked (sealed
switch cover)
> You can only
perform external
calibration by
removing this seal.
In this case,
verification is no
longer valid and
you must have your
balance/scale
re-verified.