98648-009-08 lite Sartorius Basic lite Sartorius Gem Electronic Analytical and Precision Balances and Gold Scales with EC Type Approval Installation and Operating Instructions...
Initial verification must be done at the Models BL120-0CE and BL1200-0CE: place of installation by an authorized These models are supplied for use in Sartorius representative or a local veri- legal metrological (legal-for-trade) fication officer. applications in Germany and already The balances/scales must be ad- verified ex factory.
Contents Intended Use ..........2 Initial Verification .......... 2 Contents ............3 Warning and Safety Information ...... 4 Getting Started ..........4 Operating the Balance/Scale Basic Weighing Function ......... 9 Calibration/Adjustment ........10 Application Programs Toggle between Weight Units......12 Net-Total Formulation/ Second Tare Memory ........13 Counting............14 Weighing in Percent........16...
Complete the warranty equipment. Keep these instructions registration card, and return the card to in a safe place. your Sartorius office or dealer. Follow the instructions below to Storage and Shipping Conditions ensure safe and trouble-free operation of your balance/scale:...
Page 5
Installation Instructions balance/scale. The control seal Your balance/scale is designed to consists of a sticker with the “Sartorius” provide reliable weighing results under logo. This seal will be irreparably normal ambient conditions. When damaged if you attempt to remove it.
Page 6
Setting Up the Balance/Scale Balances/Scales with an analytical draft shield chamber § Place the components listed below on the weighing chamber in the order given: – Shield plate – Shield ring – Pan support – Weighing pan/load plate – Gem tray (only for GC503) Balances/Scales with a Round Weighing Pan §...
Page 7
Connecting the Balance/Scale to AC Power/ Safety Precautions Use only original Sartorius AC adapters: – for Europe: 69 71948 – for the USA: 69 71947 – for the RSA: 69 71949 – for Australia: 69 71950 § Insert the right-angle plug into the jack §...
Page 8
Leveling the Balance/Scale Level the balance/scale any time you set it up in a new location. Use only the 2 front feet of the balance/scale for leveling. § Turn the 2 rear feet until they are in position (only for balances/scales with rectagular weighing pans/load plates) §...
Operating the Balance/Scale Basic Weighing Function Available Features – Taring the balance/scale You can tare the balance/scale within the entire weighing range. Preparation § Turn on the balance/scale: Press e § To change configurations: See the chapter entitled “Configuring the Balance/Scale” §...
Calibration/ – There is no load on Calibration/Adjustment the balance/scale Adjustment* for Verified Balances/ – The balance/scale Scales of Accuracy Always calibrate/adjust is tared Class K and m the balance/scale after – The internal signal – The external calibra- setting it up in a new is stable tion function is location.
Page 11
Example Calibrate the balance/scale Menu code settings: Factory-set codes Step Display (or instruction) 1. Turn on the balance/scale 2. To access the calibration switch: open the cover plate located below the balance/scale; flip the switch 3. Tare the balance/scale + 1000.0 4.
Application Programs All application programs can also be used in legal metrology. Calculated weight values are indicated by one of the following symbols displayed to the right of the numerical value: – Percentage – Piece count = pcs – Other calculated value = o The application programs for calibrated balances/scales not intended for use in legal metrology are labeled with a triangle # in the right side of the display (e.g.
Net-Total Formulation/Second Tare Memory With this application program you can weigh in components for formulation of a mixture. Preparation Configure the Net-total formulation/Second tare memory application in the operating menu: See “Configuring the Balance/Scale” Menu code 2.1. 3 Example Step Display (or instruction) 1.
Counting Purpose The Counting program allows you to automatically count the number of parts that have approximately the same weight. Available Features – Store the current weight value to have it loaded as the preset reference sample quantity next time you initialize the Counting application –...
Page 15
Example Determine an unknown piece count; weigh the preset reference sample quantity Settings (changes in the factory settings required for this example): Menu: Application program: Counting (2. 1. 4), Reference sample quantity: 20 pcs (3. 3. 3) Step Display (or instruction) 1.
Weighing in Percent Purpose This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight. Available Features – Store the current weight value to have it loaded as the preset reference percentage next time you initialize the Weighing in Percent application –...
Page 17
Example Determine an unknown percentage; store the weight on the balance/scale as a reference percentage Settings (changes in the factory settings required for this example): Menu: Application program: Weighing in percent ( 2. 1. 5) Menu: Reference percentage 100 % (3. 3. 5) Step Display (or instruction)
Averaging Purpose Use this program to determine weights under extreme- ly unstable ambient conditions. In this program, the balance/scale calculates the weight as the average value from a defined number of individual weighing operations. These weighing operations are also known as “subweighing operations” or “subweighs.” Available Features –...
Page 19
Example Determine the weight of a sample in extremely unstable ambient conditions by calculating the average of 10 subweighing operations. Settings (changes in the factory settings required for this example): Menu: Application program: Averaging (menu code 2. 1.12) Step Display (or instruction) 1.
Configuring the Balance/Scale Setting the Parameters (Menu Codes) You can configure your balance scale to meet individual requirements by selecting from the parameters available in the menu. Example: Adapt the balance/scale to extremely unstable ambient conditions: Menu code 1, 1. 4 Step Display (or instruction)
The balance/scale has not Calibrate/adjust the is obviously wrong been calibrated/adjusted balance/scale The balance/scale was not Tare before weighing tared before weighing The balance/scale is not level Level the balance/scale If any other errors occur, contact your local Sartorius Service Center!
Regular servicing by a Sartorius technician will extend the service life of your balance/scale and ensure its continued weighing accuracy. Sartorius can offer you service contracts, with your choice of regular maintenance intervals ranging from 1 month to 2 years.
– The AC adapter no longer functions properly – The AC adapter has been stored for a relatively long period under unfavorable conditions In this case, notify your nearest Sartorius Service Center or the International Technical Support Unit based in Goettingen, Germany. Maintenance and...
Information on EC Verification Initial Verification Performance of initial verification by Sartorius AG is documented on the balance/scale by the following label: 1. Explanation: EC mark of conformity Green “M”: Initial verification has already been performed on this instrument Year in which the initial verification...
Page 29
Using Verified Balances/Scales Approved for Use as Legal Measuring Instruments in the EU* The balance/scale is not allowed to be used for weighing goods intended for direct sale to the public. The type-approval certificate for verification applies only to non-automatic weighing instruments; for automatic operation with or without auxiliary measuring devices, you must comply with the regulations of your country applicable to the place...
Page 30
Only to be filled out by Sartorius service technicians authorized to perform initial verification* for balances/scales of accuracy class K and l in the case of verification at the site of application: Initial verification (Date): Verification is valid for the following place...
EC Directives has been extended to the Member States of the European Union and the Signatories of the Agreement on the European Economic Area. Sartorius complies with the EC Directives and European Standards in order to supply its customers with weighing instruments that feature the latest advanced technology and provide many years of trouble-free service.
Notified Body registered at the Commission of the European Community for performing such verification. The legal basis allowing Sartorius to perform EC verification is constituted by the EC Council Directive No. 90/384/EEC on non-automatic weighing instruments that has been in effect since January 1, 1993, in the Internal Market as well as...
Page 36
All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG. The status of the information, specifications and illustrations in this manual is indicated by the date given below.