Error Codes; Data Output; Saving Data - Sartorius Signum 1 Operating Instruction

Complete scales
Hide thumbs Also See for Signum 1:
Table of Contents

Advertisement

Errors
– If a key is inactive, "———-" or
"No function" is displayed briefly
(2 seconds) and an acoustic signal
(double-beep) is emitted
– Temporary errors are displayed for
2 seconds in the measured value/result
line (e.g.,
Inf 09
); fatal errors (e.g.,
Err 101
) can be cleared by switching
the scale off and then on again.
Error codes are described in detail
under "Error Codes" on page 90.

Data Output

Printer
You can connect two strip or label printers
to Signum 1, 2 or 3, and have printouts
generated at the press of a key or auto-
matically. Printout formats are user-defin-
able. You can also configure separate
summarized printouts, and print a list of
the active menu settings. See "Configuring
Printouts" on page 80 for details.
Digital Input/Output
Interface + Optional I/O
The digital I/O interface is supported
by the Checkweighing and Classification
applications.
Checkweighing
The output device has a number
of control functions. Four data outputs
transfer signals for "less than," "equal to,"
"greater" and "set." You can define
whether the outputs are always active or
are activated only at stability, only within
the checkweighing range, only within
the checkweighing range at stability, or
switched off.
Classification
Four data outputs transfer information on
the class of the load (Class 1, 2, 3, 4 or 5)
and indicate when the minimum load is
exceeded (Set). You can define whether
the outputs are always active, activated
only at stability, or off. For details,
see "Classification" on page 68.
COM Port
You can define a number of parameters
for this SBI and SMA interface (print com-
mand, time-dependent autoprint, ID
codes). See "Interface Port" on page 82 for
details.
Backup
Application parameters (such as reference
values) are saved when you change appli-
cation programs or switch off the Signum.
You can assign a password to prevent
unauthorized users from changing settings
in the "Device parameters" menu under:
Setup
Password
See also page 18.
15

Hide quick links:

Advertisement

Table of Contents
loading