Test The - User Threshold - HP 1660 Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

Test the − User threshold
Move the cursor to the field to the right of Pod 1. Type –6.00, then use the left and
1 1
right cursor control keys to highlight V. Press the Select key.
On the function generator front panel, enter − 5.718 V ± 1 mV DC offset. Use the
2 2
multimeter to verify the voltage.
The activity indicators for pod 1 should show all data channels and the J-clock channel at a
logic high.
Using the Modify down arrow on the function generator, decrease offset voltage in
3 3
1-mV increments until all activity indicators for pod 1 show the channels at a logic
low. Record the function generator voltage in the performance test record.
Using the Modify up arrow on the function generator, increase offset voltage in
4 4
1-mV increments until all activity indicators show the channels at a logic high.
Record the function generator voltage in the performance test record.
Testing Performance
To test the threshold accuracy (logic analyzer)
3–13

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660a series1660as series1660a1661a1662a1663a

Table of Contents