c c Adjust the pulse generator channel 1 Delay, then select Precision Edge Find in the
oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to
the setup time of the setup/hold combination selected, +0.0 ps or − 100 ps.
Select the clock to be tested.
4 4
a a In the Master Clock menu, select the clock field to be tested, then select the clock
edge as indicated in the table. The first time through this test, use the top clock and
edge in the following table.
Clocks
J↑
K↑
L↑
M↑
N↑
P↑
b b Connect the clock to be tested to the pulse generator channel 2 output.
c c Select Done to exit the Master Clock menu.
To test the single-clock, single-edge, state acquisition (logic analyzer)
Testing Performance
3–31