To test the time interval accuracy (logic analyzer)
Set up the Waveform menu.
4 4
a a Press the Waveform key.
b b Move the cursor to the sec/Div field, then use the RPG knob to dial in 2.00 µs.
c c Select the Markers Off field, then select Pattern.
d d Select the Specify Patterns field. Select X entering 1 and O entering 1.
e e Select Done to exit the Specify Patterns menu.
f f Move the cursor to the X-pat field. Type 1, then press Done.
g g Move the cursor to the O-pat field. Type 20, then press Done.
h h Select the Markers Patterns field, then select Statistics. Select Reset Statistics to
initialize the statistics fields.

Connect the logic analyzer

Using a 6-by-2 test connector, connect channel 0 of Pod 1 to the pulse generator
1 1
channel 1 output.
Using the SMA cable and the BNC adapter, connect the External Input of the pulse
2 2
generator to the Main Signal of the function generator.
3–62