Set Up The Logic Analyzer - HP 1660 Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

Set up the logic analyzer

Set up the Channel menu.
1 1
a a Press the Config key.
b b At the pop up menu, select Scope Channel.
c c Select the Input field, then select C1.
d d Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
e e Move the cursor to the V/Div field, then use the RPG knob to dial in 20 mV.
f f Move the cursor to the Offset field. Set the Offset to 0 V by typing 0, then pressing the
Select key.
g g Select the Coupling field, then select 50Ω / DC.
Set up the Trigger menu.
2 2
a a Press the Trigger key.
b b Select the Mode/Arm field, then select Immediate.
To test the input resistance (oscilloscope)
3–69