Set Up The Logic Analyzer - HP 1660 Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

To test the CAL OUTPUT ports (oscilloscope)

Set up the logic analyzer

Set up the Calibration menu.
1 1
a a Press the Waveform key.
b b Press the Waveform key again. At the pop up, select Scope Calibration.
c c Select the Mode field, then select Service Cal.
d d Select the Procedure field, then select DC Cal BNC.
e e Select the DC volts field, and set it to O V.
3–65