Acquire the data

Enable the pulse generator channel 1 output (with the LED off).
1 1
Press the blue key, then press the Run key to select Run-Repetitive. Allow the logic
2 2
analyzer to acquire data for at least 100 valid runs as indicated in the pattern
statistics field.
When the logic analyzer has acquired at least 100 valid runs, touch Stop. The
3 3
Min X-O field in the logic analyzer Pattern Statistics menu should read
94.99–95.00 µs. The Max X-O field should read 95.00–95.01 µs. The Avg X-O field
should read 94.99–95.01 µs. Record the results in the performance test record.
3–63