Testing Performance
To test the single-clock, single-edge, state acquisition (logic analyzer)
Set up the oscilloscope.
3 3
Oscilloscope Setup
Time Base
Time/Div: 1.00 ns/div
Channel
Display
Probe Atten
Offset
Volts/Div

Set up the logic analyzer

Set up the Configuration menu.
1 1
a a Press the Config key.
b b In the Configuration menu, assign all pods to Machine 1. To assign the pods, select
the pod fields, then select Machine 1 in the pop-up menu.
c c Select the Type field in the Analyzer 1 box, then select State.
3–24
Display
Delta V
avg
V markers on
# of avg: 16
marker 1 position: Chan 1
screen: dual
marker 2 position: Chan 1
Channel 1
Channel 2
on
on
20.00
20.00
− 1.3 V
− 1.3 V
400 mV
400 mV
Delta T
T markers on
start on: Pos Edge 1
stop on: Neg Edge 1